③ Use the diode test mode to check diodes, transistors and other
semiconductor device. In the diode test mode sends a current
through the semiconductor junction, and the measure the
voltage drop across the junction. A good silicon junction drop
④ For forward voltage drop reading on any semiconductor
component, place the red test lead on the component anode
and place the black test lead on the component cathode. The
measured value show on the display.
⑤ Reverse the test lead and measure the voltage across the diode
●
If diode is good, the display shows “OL”.
●
If diode is shorted, the display shows 0 (zero) in both
●
If display shows “OL” in both direction, the diode is open.
① Press the “FUNC” key to enter to the continuity mode.
② The buzzer sound if the resistance of a circuit under test is less
5-4. Frequency measurement [ Only for ADM01]
① Set the rotary switch to “Hz” position.
② Insert the red test lead into the “ VΩHz” input terminal and the
black test lead into the “COM” terminal.
③ Connect the test leads across with the circuit under testing. The
measured value shown.on the LCD display.
NOTE: Input signal level must be higher than 0.5V (it is sensitivity).
5-5. DC/AC µA or mA measurement
DC Current range is 200.0µA/2000µA and 20.00mA,/200.0mA and
AC Current range is 200.0µA/2000µA and 20.00mA,/200.0mA and
① Turn off power to the circuit. Set the rotary switch to the