Operating Manual 39
Starting Up
6. Enable remote communication
Action How to do it Picture
1
Install TCPIP or
phone wiring from
the SCM controller
to your server/
computer.
Plug an Ethernet or phone line into the back of the SCM controller, then make sure the wire is
connected to your server or computer.
2
Set the network
address.
a. From the Communications window, click
System Configuration.
b. In the System Configuration
window, click on the Network
Settings tab.
c. Enter the following network
settings as determined by your
system administrators:
Address: TCPIP address
assigned to the SCM.
Example: 192.168.1.246
Subnet Mask: Example:
255.255.255.0
Gateway Address: Example:
192.168.1.1
Alarm Destination Address:
TCPIP address of the computer
which is to be automatically
notified of alarms by the BVS
Observer program. Example:
192.168.1.54
Alarm Destination Port: Port
address of the computer which
is to be automatically notified
of alarms, as determined by your system administrator. Example: 8842
d. Click OK when done.