44 S5 Battery Validation System
8
Identify where the
temperature sensors
are located.
a. In the Software Limit
Settings window, in the
Apply to Unit(s) field,
select a unit that has a
thermistor attached, then in
the Temperature Sensor
field, select the ordinal
number (e.g. 1st, 2nd, 3rd,
etc.) of the thermistor that
goes with it.
b. Click OK when done.
9
Enter the connection
information.
a. In the Software Connection Settings window, enter the following data:
• To use a serial line: Leave the Phone Number field blank and choose 19,200 in the Speed field.
• To use a phone line: Enter the phone number of the SCM in the Phone Number field.
• To use a TCPIP connection: Enter the network IP address in the Network Address field and
selelct the network port number in the Network Port field.
b. Click OK when done.
10
Add the battery file
you just created.
1 In the Location List window, click on Add Location to List.
2 When prompted, browse for the new battery location file. Look in the directory you specified
in step 3. The default location for battery location files is C:\BVM4Data\.
3 Click on Open when done.
Action How to do it Picture
Serial line
Phone line TCPIP Connection