26
Melting point B-540
Modifications in the ready mode
Modifications between SETPOINT and MAXPOINT
Problem: The temperature GRADIENT and/or MAXPOINT have
been selected too low.
The field for temperature GRADIENT begins to flash and can
be increased via UP/DOWN.
The new temperature GRADIENT is acknowledged with
ENTER and the MAXPOINT temperature flashes.
The MAXPOINT temperature can be increased via UP/DOWN.
The newly selected MAXPOINT is confirmed with ENTER and
the unit returns to the determination mode.
Problem: The determination is to be aborted.
The determination is aborted and the user returns to the main
menu.
5 Operation
PRESS START
TEMPERATURE 135.6 °C
GR 2.0 °C
MP 145 °C
TEMPERATURE : 140 °C
GR 2.0 °C
MP 145 °C