Melting point B-5409 Appendix
Dimensions of the unit (width x height x depth) 170 x 210 x 340 mm
Dimensions of packaging (width x height x depth) 340 x 390 x 580 mm
Weight 5.4 kg
Temperature range 25 °C - 400°C
Cooling-down rate 350°C – 50°C max. 10 min
Heating-up speed 50°C – 350° C max. 7 min
Measuring accuracy up to 100°C +/- 0.3°C
Measuring accuracy up to 250°C +/- 0.5°C
Measuring accuracy up to 400°C +/- 0.8°C
Temperature ramps 0.1/0.2/0.5/1.0/2.0/5.0/10.0/20.0°C min.
Mains voltage 100 – 120 V/ 50 or 60 Hz
220 – 230 V/50 or 60 Hz
Power consumption approx. 200 W
Current consumption approx. 2 A (100-120 V) / 1 A (220-230 V)
Data saving with current failure The status of the unit is stored
Restarting is required
Ambient temperature 5 – 40°C
Table 6: Technical data
Fault Cause Remedy
STORED DATA LOST The battery is exhausted Press the Menu key for 5 seconds.
PLEASE CHECK BATTERIE Order new battery and have it installed by techn.
AND SET TIME AND DATE service. Date and time to be reprogrammed.
CALIBRATION INTERVAL The selected calibration Press the Menu key for 5 seconds.
EXCEEDED! time has expired. Calibrate the unit.
LAST CAL: XX:XX (See chapter 6, Maintenance)
The Citizen printer does Configuration The SEL key on the printer must
not print be actuated whenever he printer is switched on.
CALLSERVICE Damage in the unit Switch off the unit (main switch).
ERROR 001/099 Press START-key.
Switch on the unit and keep holding the START-key.
Wrong melting points. Wrong transmission. Always remove capillaries when
switching the unit on.
Calibration Calibrate the unit.
Table 5: Error messages
9.7 Technical data
9.6 Error messages