Melting point B-540
5.7.3 Modifications of parameters during a
melting point determination
Quite frequently, the user will find out during an actual
determination that his selected parameters have to be
adapted, e.g. because the SETPOINT or the MAXPOINT are
too low or because the GRADIENT runs too slowly. In order
to adapt these parameters, it is not necessary to start the
determination again from the beginning.
The following modifications are possible:
Modifications before reaching the SETPOINT
The SETPOINT can be adapted.
The determination can be aborted.
Cooling down to ambient temperature.
Modifications in the ready mode
The MAXPOINT can be adapted.
The GRADIENT can be adapted.
The determination can be aborted.
Cooling down to ambient temperature.
Modifications between SETPOINT and MAXPOINT
The MAXPOINT can be adapted.
The GRADIENT can be adapted.
The determination can be aborted.
Cooling down to ambient temperature.
Modifications before reaching the SETPOINT
Problem: The SETPOINT has been selected too low.
The field for SETPOINT temperature begins to flash and may
be rised by actuating UP/DOWN.
The new SETPOINT temperature is confirmed with ENTER
and the unit returns to the determination mode.
Problem: The determination is to be aborted.
The determination is to be aborted and the unit is to be cooled
down to ambient temperature.
5 Operation
TEMPERATURE 25.0 °C
SP 136 °C