8000302 | Instructions for use RUCK Podiatry Couch MOON | REV 04 |
Revision status 08/09/2023
31
Set-up
4.10 Connect to the mains power supply
Prior to start-up, check whether the mains voltage matches the rated voltage
and frequency specied on the rating plate (on the rear of the sheet metal
cladding). Connect the podiatry couch with the provided mains cable (rolled
cable length: 180cm) to a power socket with a protective earth contact.
The mains plug and socket must be easily accessible so that the podiatry couch
can be quickly and easily disconnected from the mains supply. Alternatively,
you can connect the couch to a socket that can be disconnected from the
mains by a switch.
If delivered at a cold time of the year (outdoor temperature <5°C), a condition-
ing period of at least 48h should be observed before the podiatry couch may
be turned on.
WARNING
Lay the mains cable and the foot switch cable such that the cables
cannot get in the way of moving parts of the couch. Crushed cables
can make the couch electrically live or cause a short circuit. Risk of
fatal injury from electric shock!
Lay all cables such that no cable can create a tripping hazard. Risk of
injury!
Make sure the couch can be quickly disconnected from the mains
power in an emergency. Make sure the mains plug is easily accessi-
ble or use a socket that can be switched o.