Technicalspecications
Test Limit Electromagnetic environment –
guidelines
Line-conducted
emissions
CISPR 11, Group 1,
Class B
The device uses RF energy for internal
functions only. Accordingly, it only
gives o very few RF emissions, which
probably cause no interference with
nearby electronic devices.
Interference radi-
ation
CISPR 11, Group 1,
Class B
emissions IEC 61000-3-2,
Class A
The device is suitable for use in envi-
ronments that are directly connect-
ed to the public low-voltage power
network.
Voltageuctua-
tions/icker
IEC 61000-3-2,
Class A
The device is suitable for use in envi-
ronments that are directly connect-
ed to the public low-voltage power
network.
Test Compliance level Electromagnetic environment –
guidelines
Electrostatic dis-
charge
(IEC 61000-4-2)
Contact discharge:
± 8 kV
Air discharge: ±2 kV,
±4 kV, ±8 kV, ±15 kV
The oor covering should be made of
wood, concrete or ceramic tiles. If the
oor covering is a synthetic material,
the relative air humidity should be at
least 30%.
7.2 Electromagnetic interference emission
7.3 Electromagnetic immunity