Technicalspecications
Test Compliance level Electromagnetic environment –
guidelines
Radiated RF EM
eld
(IEC 61000-4-3)
80-6000 MHz; 1
kHz AM 80%; 3 V/m
Portable and mobile radio devices shall
not be used within a closer proximity
to the (device or system), including
the cables, than the recommended
safety distance, calculated using the
equation applicable for the transmis-
sion frequency.
Recommended safety distance
d = 1.2√P for 150kHz to 80MHz
Where "P" is the nominal power of the
transmitter in watts (W) according
to the information provided by the
transmitter manufacturer, and "d" is
the recommended safety distance in
meters (m).
Proximityelds
of RF wireless com-
munication devices
(IEC 61000-4-3)
385 MHz; pulse
modulation:
18 Hz; 27 V/m
450 MHz; pulse
modulation:
18 Hz; 28 V/m
710, 745, 780 MHz;
Pulse modulation:
217 Hz; 9 V/m
810, 870, 930 MHz;
Pulse modulation:
18 Hz; 28 V/m
1720, 1845, 1970
MHz;
Pulse modulation:
217 Hz; 28 V/m
2450 MHz; pulse
modulation:
217 Hz; 28 V/m;
5240, 5500, 5785
MHz;
Pulse modulation:
217 Hz; 9 V/m
Portable and mobile RF communica-
tion devices should not be used within
a distance of 30cm to any part of the
device, including the cable.