EasyManua.ls Logo

Buck MOON - Page 61

Buck MOON
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8000302 | Instructions for use RUCK Podiatry Couch MOON | REV 04 |
Revision status 08/09/2023
61
Technicalspecications
Test Compliance level Electromagnetic environment –
guidelines
Interference
immunity to rapid
transient electrical
interference var-
iables/burst (IEC
61000-4-4)
DC power supply
line: ± 2 kV; 100 kHz
repetition frequency
Signal line: ±1kV;
100 kHz repetition
frequency
Voltage spikes
(surges)
(IEC 61000-4-5)
L-N: 1 kV at
0°,90°,180°,270°
L-PE, N-PE: 2 kV at
0°,90°,180°,270°
Line-conduct-
ed disturbance
variables, induced
by high-frequency
elds(IEC61000-
4-6)
0.15–80 MHz; 1 kHz
AM 80%; 3 VRMS
and 6 VRMS in ISM.
Portable and mobile radio devices shall
not be used within a closer proximity
to the (device or system), including
the cables, than the recommended
safety distance, calculated using the
equation applicable for the transmis-
sion frequency.
Recommended safety distance
d = 1.2√P for 150kHz to 80MHz
Where "P" is the nominal power of the
transmitter in watts (W) according
to the information provided by the
transmitter manufacturer, and "d" is
the recommended safety distance in
meters (m).
Power-frequency
magneticelds
(IEC 61000-4-8)
30 A/m, 50 Hz