LySonix 3000
®
with PulseSelect™ Operation Manual
40-OM-0002 Page 35 of 43 Rev B
8. SYSTEM DISASSEMBLY AND MAINTENANCE
8.1 Cleaning and Sterilization
CAUTION - All Probes, Ha
ndpieces, Handpiece Cables, Wrenches and accessories are supplied
industrially clean, but non-
sterile. All items intended for use in the sterile field must be cleaned and
sterilized as per the following instructions before first use and before every subsequent use.
8.1.1 Follow ANSI/AAMI ST35, Good Hospital Practice: Handling and Biological
Decontamination of Reusable Medical Devices (1991) or latest edition, or such other
guidelines as may be directed by Hospital or Clinic GHP’s.
8.1.2 Disassemble all components of system in reverse order of assembly (refer to Figure 2),
including the Probe from the Handpiece, the Front Cap from the Handpiece and the tubing
from the Handpiece, and the Irrigation and/or Aspiration Unit. Unplug all power cords.
Use Wrenches supplied and rotate wrenches as shown in Figure 3 to untighten Probes from
Handpiece. Separate Probes from Handpiece.
8.1.3 Dispose of all flexible tubing (but NOT the Handpiece Cable if detachable) and the Waste
Reservoir Container in accordance with standard hospital procedures for disposal of
contaminated biological wastes.
8.1.4 Rinse the Probes, Wrenches and Handpiece under warm running water. Handpiece should
not be immersed fully under water.
8.1.5 Brush hole in Handpiece and Hollow Probes using the brushes as follows:
Item To Be Cleaned Brush
3.5mm Diameter Probe Small Cleaning Brush-Probe
4mm Diameter Probe Small Cleaning Brush-Probe
5mm Diameter Probe Small Cleaning Brush-Probe
Handpiece Large Cleaning Brush-Handpiece
8.1.6 Brushing should be done with anti-microbial and anti-viral soap and warm running water.
This insures clearing of debris from the internal passages. Rotate brush as it is inserted to
help clear debris and stains. Brushes should be inserted fully through the unit. Repeat this
procedure at least four (4) times. A standard soft bristle cleaning brush may be used to
scrub exterior surfaces of Handpiece, Front Cap and Probes. Rinse all soap residue from
interior and exterior of unit under warm running water for a minimum of 1 minute.
Handpiece should not be immersed fully under liquid.