LySonix 3000
®
with PulseSelect™ Operation Manual
40-OM-0002 Page 37 of 43 Rev B
8.1.12.3 Cable:
121°C gravity 50 minutes wrapped or unwrapped
132°C gravity 10 minutes unwrapped, 20 minutes wrapped
PREVAC, 132°C 3 minutes wrapped or unwrapped
8.1.13
Dry cycle: MINIMUM of 5 minutes is recommended.
8.2 Reuse Life of Components
8.2.1 The estimated reuse life of system components is listed below. All reuse life estimates are
approximate and may be affected by rough handling, damage, and wear due to vigorous
cleaning, etc.
Component Reuse Life (Number of sterilization cycles)
Handpiece 200 cycles
Handpiece Cable 200 cycles
Front Cap 200 cycles
All Probes >200 cycles
Wrenches >200 cycles
NOTE: -
The Reuse Life given takes into account wear and tear due to cleaning and sterilization only.
Damage or wear caused by actual use in surgery will affect life of components.
8.3 Deviations from Cleaning and Sterilization Instructions
8.3.1 It is highly recommended that the procedures given in this manual for cleaning and
sterilizing the LySonix 3000
®
with PulseSelect
and related accessories be followed. It is
the responsibility of the user of this device or any accessories used with it to validate
procedures for cleaning and/or sterilization if they differ from the procedures as outlined in
this manual.