Oxygen Concentrator Technical Service Manual | PN 20631679 Rev G
8
Air Inlet, HEPA, Compressor Intake
IEC Class I, Type B Applied Part, IP0 (3 LED model)
IEC Class II, Type BF Applied Part, IP22 (2 LED model)
Pulse Dose Mode Specifications
1.0 to 6.0: 16mL, 24mL, 32mL, 40mL, 48mL, 56mL, 64mL, 72mL, 80mL, 88mL, 96mL;
7-9: 128mL, 160mL, 192mL
Adjustable between settings of 1 (most sensitive) to 3 (least sensitive)
Adjustable Bolus Rise Time
Adjustable settings of Fast(factory setting), Medium, or Slow
• Cannula pressure has dropped below the trigger point (typically between 0.135—0.37 cm of H
2
O
of negative pressure)
• At least 1¼ seconds has passed since the last pulse began
Minimum time between breaths
1.25 seconds (max. 3 consecutive breaths)
Response to Missing Breaths
While in Pulse Dose Mode, the Eclipse 5 is always monitoring for breath detection. After 15 seconds of no
breath detected, the Eclipse 5 “delivers” Continuous Flow at the last Continuous Flow setting.
After another 15 seconds, the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a
breath. The Eclipse 5 will stay in this modality until a breath is detected. (See Power Cartridge
Conservation Feature, page 26.)
The expected service life is a minimum of 5 years.
AC Power Supply and Power
Cartridge (Battery)
NOTE: Bolus volume decreases as breath rate exceeds published range.
Power Accessory Specifications
AC Power Supply
DC Power Supply
Power Cartridge (Battery)
Quantity (2) 97.5 W-hrs batteries
(Each containing 7.92 grams equivalent Lithium content)
Nominal Power
Cartridge Life
80% Capacity after 500 Charge/Discharge cycles
Power
Cartridge
Recharge Time
1.8 to 5.0 hours, dependent on flow setting, to achieve 80%
capacity from a fully discharged Power Cartridge