ELECTROMAGNETIC COMPATIBILITY
This equipment has been tested and found to comply with the limits for medical devices to the IEC60601-1-2 Electromagnetic
Compatibility standard. These limits are designed to provide reasonable protection against harmful interference in a typical
medical installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed according to the
instructions, may cause harmful interference to other devices in the vicinity. There is, however no guarantee that interference will
not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined
by turning the equipment off and on, the User is encouraged to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving device.
• Increase the separation distance between the equipment.
• Connect the equipment into an outlet on a circuit different from that which the other device(s) are connected.
• Consult with CAIRE’s Technical Support Department for help.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according
to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
The use of Accessories, transducers, and cables other than those specified, with the exception of transducers and cables sold by
the Manufacturer of this device as replacement parts for internal components, may result in increased Emissions or decreased
Immunity of the Eclipse 5.
The Eclipse 5 should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the
Eclipse 5 should be observed to verify normal operation in the configuration in which it will be used.