EasyManua.ls Logo

CAIRE Hi Flow Stroller - Page 8

CAIRE Hi Flow Stroller
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
Hi Flow Stroller Technical Service Manual • PN 14314465 Rev F
7
Hi Flow Stroller
Safety
V
Table 2
Guidance and manufacturers declaration—electromagnetic immunity
The Hi Flow is intended for use in the electromagnetic environment specified below. The customer or the user of the Hi Flow should
assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment—guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or ceramic tile. If floors
discharge (ESD) ±8 kV air ±8 kV air are covered with synthetic material, the relative
IEC 61000-4-2 humidity should be at least 30%.*
Electrical fast ±2 kV for power Not applicable Not applicable
transient/burst supply lines DC powered device
IEC 610004-4 ±1 kV for Not applicable
input/output lines No data input/output lines
±1 kV line(s)
Surge to line(s) Not Applicable Not Applicable
IEC 61000-4-5 ±2 kV line(s) DC powered device
to earth
Voltage dips, <5% UT (>95% dip
short interruptions in UT) for 0,5 cycle
and voltage 40% UT (60% dip
variations on in UT) for 5 cycles Not Applicable Not Applicable
power supply 70% UT (30% dip DC powered device
input lines in UT) for 25 cycles
IEC 61000-4-11 <5% UT (>95% dip
in UT) for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels
(50/60 Hz) characteristic of a typical location in a typical
magnetic field commercial or hospital environment.
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
* This statement indicates that the required testing was performed in a controlled environment and the Hi Flows are found to be compli-