EZ-IN
®
Series Turbine Flowmeter Manual
14
Maintenance and Repairs
If owmeter readings appear erratic, check the owmeter for wear or debris.
Under normal operation with non-abrasive uids, owmeters can provide
years of service without repairs. If a owmeter is exposed to abrasive uids
or long periods of over-range ow rates, an annual check is recommended.
Removing the Meter from the Pipeline
1. Remove all pressure from the line.
Remove the magnetic pickup (Figure 5, Page 13) to avoid damaging it.
Remove one or more of the ange bolts.
Loosen the remaining ange bolt nuts to allow clearance for the meter.
If additional “spreader” nuts were installed, tighten them to help free
the meter.
Lift the meter from between the pipeline anges, using an installation
handle if necessary (Figure 3, page 8).