Appendix F.  Time Series and Spectrum 
Graph Information 
The AVW200 uses an audio A/D for capturing the sensor’s signal.  The 
number of samples acquired in this period is 4096 points.  A Fast Fourier 
Transform (FFT) algorithm is used to create a frequency spectrum.  The 
frequency spectrum is displayed in the graph labeled “Spectrum” (see Figure 
1.1-1).  This graph shows each of the frequencies and the voltage amplitude in 
mV RMS. 
The “Time Series” graph is the acquired or sampled data in the time domain. 
The graph shows the combination of all the frequencies coming from the 
vibrating wire sensor shortly after the sensors excitation. The dominate 
frequency is the natural resonating frequency of the vibrating wire. The other 
frequencies can include noise pickup (i.e., motors close to the sensor, pickup 
from long wires), harmonics of the natural frequency or harmonics of the noise 
(50/60 Hz harmonics) and/or mechanical obstruction (loosing of the wire or the 
wire vibration is physically changed by the package movement).  The 
AVW200 computes a signal-to-noise diagnostic by dividing the response 
amplitude by the noise amplitude. 
The “Time Series” graph shows the decay from the start of the sampling to the 
end of the sampling.  The decay is the dampening of the wire over time.  The 
AVW200 computes a decay ratio diagnostic from the time series ending 
amplitude divided by the beginning amplitude.  Some sensors will decay very 
rapidly, others not.  It is a good idea to characterize the sensors decay and 
amplitude when the sensor is new, so that over time the health of the sensor can 
be monitored. 
By changing the begin and end frequencies in the options tab, the affects of 
narrowing can be of value for troubleshooting and solving problems with errant 
sensors, or improving the measurement.  Care should be taken to ensure that 
when you change the begin and end frequency that the frequency range still 
captures the sensor’s signal. 
F.1  Good Sensor Examples 
Figure F.1-1 and Figure F.1-2 are measurement results from the same sensor — 
the first measurement was taken with a swept frequency between 200 and  
2200 Hz while the second measurement was taken with a swept frequency 
between 200 and 6500 Hz.  Using the tighter frequency range (Figure F.1-1), 
the measurement recorded the greatest sensor noise at a frequency of 935 Hz 
with a signal-to-noise ratio of 318.  Sweeping the same sensor over the far 
wider range of 200 to 6500 Hz (Figure F.1-2) uncovers noise at 4150 Hz with a 
signal-to-noise ratio of 21.4, which is 15 times less than the signal-to-noise 
ratio of the first measurement.  This illustrates that better readings are produced 
when the sensor is swept over more narrow frequency ranges.  Also, with the 
narrowed range (Figure F.1-1), the noise frequency that exists at 4150 Hz is 
completely ignored and is not relevant because it lies outside the sampling 
frequency range; excitation is limited outside the swept frequency range as 
well. 
F-1