Appendix H.  Additional Programming Examples 
 
 Scan (2,Min,0,0) 
  PanelTemp (PTemp,250) 
  Battery (Batt_volt) 
 
  AVW200(VWResults(1),ComME,0,20,Mux1(1,1),1,1,16,1000,2500,2,_60Hz,1,0) 
    For i = 1 To 16 
   Amp1(i)   = Mux1(i,2) 
   Therm1(i)  = Mux1(i,6) 
   VWFreq1(i)  = Mux1(i,1) 
   Sig2Noise1(i) = Mux1(i,3) 
   DecayRatio1(i) = Mux1(i,5) 
   FreqOfNoise1(i) = Mux1(i,4) 
   Digits = (VWFreq1(i)/1000)^2 * 1000 'Convert frequency to Digits 
 
'Convert resistance to temp F. 
      Temp1(i) = (1/(A + B*LN(Therm1(i)) + C*(LN(Therm1(i)))^3)-273.15)*1.8+32  
 
'Calculate displacement (inches) from Digits and calibration polynomial 
   Displacement1(i)=PolyCoef1(3*i-2)*Digits^2 + PolyCoef1(3*i-1)*Digits + PolyCoef2(3*i) 
 
  Next i 
 
'Sensors 1-8 are excited over the freq range of 1000 - 2500 
AVW200(VWResults(2),ComME,0,20,Mux2(1,1),2,1,8,1000,2500,2,_60Hz,1,0) 
 
'Sensors 9-16 are excited over the freq range of 450 – 6500 
  AVW200(VWResults(2),ComME,0,20,Mux2(9,1),2,9,8,450,6500,2,_60Hz,1,0)  
    For i = 1 To 16 
   Amp2(i)   = Mux1(i,2) 
   Therm2(i)  = Mux1(i,6) 
   VWFreq2(i)  = Mux1(i,1) 
   Sig2Noise2(i) = Mux1(i,3) 
   DecayRatio2(i) = Mux1(i,5) 
   FreqOfNoise2(i) = Mux1(i,4) 
   Digits = (VWFreq2(i)/1000)^2 * 1000 'Convert frequency to Digits 
 
'Calculate displacement (inches) from Digits and calibration polynomial 
   Displacement2(i)=PolyCoef2(3*i-2)*Digits^2 + PolyCoef2(3*i-1)*Digits + PolyCoef2(3*i) 
 
'Convert resistance to temp F. 
      Temp2(i) = (1/(A + B*LN(Therm2(i)) + C*(LN(Therm2(i)))^3)-273.15)*1.8+32  
  Next i 
 
  CallTable MuxExample 
 NextScan 
EndProg 
 
H-6