3.7 Test function of the fan
If required change fan:
- Loosen screws of the cover.
- Take filter of the fastening pin
- Loosen plug
- Pin and fasten on new fan
3.8 Check the Door(s)
- retighten all screws (door handle, locking tab)
- check the locking unit (retighten screws)
- check the hinges (retighten screws with a screwdriver PH2)
- check the sealing lips
- check also backside, if two-sided version