Caution!
If the door is open for a longer time, this will lead to increase of the humidity in the cabinet,
which will lead to longer drying times.
It is also possible to change error air pressure malfunction (relative). Factory settings are adjusted to
50%. This means if an endoscope has more than 50% of air pressure malfunction, it will be displayed
with status . Then it is recommended to reprocess the endoscope again.
Caution!
A change of error air pressure malfunction (relative) could lead to displaying a successfully
completed process, even if endoscopes have had insufficient air pressure.
4.4 Service settings
After successful completion of maintenance the service technician has to enter the next inspection date.
The next maintenance should be performed after a maximum of 12 month. Over service settings it is
also possible to change admin- and service- password.
Caution!
A change of admin- and service- password could cause problems.
Only staff with knowledge of the password can change service settings.