page 25 of 28
5 Technical Data
W: 60
H: 190
D: 54
5 °C to 40 °C
10 % to 75 %
100-240 V AC ±10%, at 50/60 Hz
Rubber connector C14 according IEC/EN 60320-1
T4A Internal device:
Building: 16 A
100 W
Max. 100 W
Max. 342 Btu/h
145
Temperature:
rel. humidity (air):
Temperature:
rel. humidity (air):
Ethernet
Pressure:
Demand:
Pressure one-click:
Demand one-click :
Diameter (DN):
Quality:
Class:
5 Bar
1x RJ45 (10/100 Mbit)
2 - 10 Bar (29 - 145 PSI)
100 l/min (ANR) ,
17 lmin @ 5 Bar
3.5 SCFM,
0.6 CFM @ 72.5 PSI
5-10 Bar (72.5 - 145 PSI)
200 l/min (ANR)
33 l/min @ 5 Bar
7 SCFM
1.2 CFM @ 72.5 PSI
6 (1/4”)
according to ISO 8573-1:2010
1.4.1
Dimensions [cm]:
Weight [kg]:
Operating conditions:
Storing conditions:
Maximum relative humidity 80% for
temperatures up to 31°C
with linear decrease to 50% of
relative humidity at 40 ° C.
Interfaces:
Compressed air connection:
Power supply voltage:
Connection:
Fuses:
Power input:
Heat emission:
-10°C to 60°C
10 % to 90 %
1 kWh /Tag
II
IK6
Average electrical consumption:
Overvoltage cathegory:
Degree:
ENDODRY™ and endoSTORE™ cabinets must be installed in an environment with the following
characteristics of temperature, humidity, pressure and a maximum installation altitude of 2000
meters above sea level.
They are designed for indoor use.