general information:
o clinic, city, department
o contact person, telephone
o possibly: order number of the maintenance or service visit
o date, name of the service technician
pneumatic tests:
o replace filter (pneumatic module)
o set the pressure control
lower threshold value: 2 bar possibly: note and write down the different value
upper threshold value: 10 bar possibly: note and write down the different value
o check warning: air pressure low/high
o check process settings drying phase 1
o check process settings drying phase 2
check process settings storage phase
possibly: note the different value
cabinet tests:
o function of the drawers (drawer rails)
o tube connections
o functionality & condition of inlet air pins
o o-rings of the nozzles (if applicable: replace)
o air outlet of all nozzles
possibly: note different values and taken measures
o replace filters of fans
o check function of fans
o check the door(s)
retighten screws (door handle, hinges, locking unit)
check hinges
sealing lips (possibly: replace)
o bar code reader (cable, plug, function)
o SD card (just for WinCE models) (function, free memory capacity)
o touch display (errors, possibly: calibrate)
o Software update if necessary, new version:
6 Checklist for maintenance of the cabinet
information of the cabinet:
o device model
o serial number
o date of installation
o software version
o last maintenance