SETUP AND CALIBRATION, Cont.
Four Percent Zero Tracking Range Limit
The display will next indicate
trL=
. This is the prompt to select whether a 4% limit be placed
on the zero tracking and the push button zero. Press the
ASTERISK
key to show current
setting, then press the UNITS key to select 1 (1=YES, enable) or 0 (0=NO, disable). If enabled
(1), the zero tracking and the push button zero is limited to 4 % of the scale capacity. If
disabled (0), zero tracking and the push button zero will operate up to the full capacity of the
scale. Press the
PRINT
key to save the setting.
Motion (Unstable) Range
The display will now show
Un5=
. Press the
ASTERISK
key to show the current setting.
Changes in weight exceeding the selected number of divisions will cause the STABLE weight
annunciator to turn off. Values from 1 to 9 divisions may be selected by pressing the UNITS
key. Once the correct value is shown press the
PRINT
key to save the setting.
Digital Filter Level Selection
The display will now show FLt= which is the prompt for the selection of the digital filtering level.
Your 204 will arrive with the factory filter setting (1=minimal) already entered. Please check
with your scale service technician should you wish to change the programmed filter level and
break range. Four levels of filtering are available. They are as follows:
0 = NO FILTERING 2 = MODERATE FILTERING
1 = MINIMAL FILTERING 3 = CUSTOM FILTERING
NOTE: Selection 3, Custom Filtering is used when 0, 1 or 2 are inadequate.
Press the
ASTERISK
key to show the current setting. Then press the
UNITS
key to select the
desired level of filtering. Press the
PRINT
key to save the setting.
F=
- Filter Level
If you select Custom Filtering, the 204 will display
F=
. Press the
ASTERISK
key to show the
current setting for the Filter Level. The filter level is a number from 1 to 16 that corresponds
to the level of filtering with 16 being the greatest filtering and 1 the least. Use the
UNITS
and
ASTERISK
keys to select the filter level, then press the
PRINT
key to save the setting.
br=
- Break Range
Next, the 204 will display
br=
. Press the
ASTERISK
key to show the current setting for the
Break Range. The break range is a number from 1 to 64 that corresponds to the number of
division change to break out of filtering. Use the
UNITS
and
ASTERISK
keys to select the
break range value, then press the
PRINT
key to save the setting.
Sample Rate
The display will now show
Sr=
. Press the
ASTERISK
key to show the current setting. The
sample rate may be set from a minimum of 1 sample per second to a maximum of 10 samples
per second in one sample per second intervals. Press the
UNITS
key until the desired sample
rate is displayed. Press the
PRINT
key to save the setting.
Hold Key Enable
The display will next indicate HLd= which is the prompt for the selection to enable or disable
the Hold (ASTERISK) key. This mode of operation will lock the 204 display and is used only in
noncommercial applications. Press the
ASTERISK
key to show the current setting, then press
the
UNITS
key to select 1 (1=YES, enable) or 0 (0=NO, disable). Press the
PRINT
key to save
the setting.
Note! This feature must be set to 0 (0=NO) for “Legal for Trade” applications.