8200-M129-O1 Rev D 205 Installation & Technical 26
SETUP AND CALIBRATION, CONT.
dPP= (Decimal Point Setting)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW key to toggle
to a new setting and then press the TARE/ENTER key to save it. Allowable values are: 0, 1, 2
or 3.
0 = XXXXXX 2 = XXXX.XX
1= XXXXX.X 3 = XXX.XXX
CAP= (Capacity)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW and
UNITS/LEFT ARROW keys to enter a new setting and then press the TARE/ENTER key to
save it. Allowable values are: 1 through 999,999. NOTE: Capacity cannot exceed 999,999.
Unit2= (Weighing Unit 2)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW key to toggle
to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
0 = none 4 = oz (ounces)
1= tn (tons) 5 = kg (kilograms)
2= g (grams) 6 = tonnes (metric tons)
3 = lb (pounds) 7 = lb/oz (pounds/ounces)
NOTE: The selection for Unit2 can not be the same as Unit1. In addition,
dependent upon the selection for Unit1 and the interval and decimal point settings,
not all unit combinations are available.
trA= (Zero Tracking Range)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW and
UNITS/LEFT ARROW keys to enter a new setting and then press the TARE/ENTER key to
save it. Allowable values are: 0 (disables Zero Tracking), .5, or 1 through 9.
trL= (4% Zero Range)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW key to toggle
to a new setting and then press the TARE/ENTER key to save it.
trL=1
Yes
trL=0
No
4% of scale capacity Full capacity (no limit)
PUO= (Power-Up Zero Feature)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW key to toggle
to a new setting and then press the TARE/ENTER key to save it.
PUO=1
Yes
PUO=0
No
Automatic Re-Zero on Power-Up No Re-Zero on Power-Up