8200-M129-O1 Rev D 205 Installation & Technical 43
SETUP AND CALIBRATION, CONT.
Weight On Demand
If continuous output has not been selected for Serial Port 0 (Cont0=NO), the 205 indicator will
respond to a weight request (ENQ).
The host device (computer) sends:
ENQ - (hex 05)
The 205 will respond:
<s><xxxxxx><d><uu><m><cc><cr>
Where:
s = Sign "-" = negative, " " (blank) = positive
xxxxxx.xxx = Weight Six digits
d = Decimal point Added to string if enabled in setup
uu = Units tn, lb, l/o, oz, t, kg, g
m = Mode G = Gross, N = Net
cc = Weight Status OC = overcap
CZ = center of zero
MO = motion
ee = weight not currently being displayed
cr = Carriage Return (hex 0D)
Sio 1? (Setup Serial Port 1)
Press the TARE/ENTER key. The display will change to show the current setting no. To skip
configuring the Sio Port 1 (serial input/output) and proceed to the Sio 2? menu, press
the TARE/ENTER key again. To configure the Sio 1, press the /UP ARROW key to select
YES and then press the TARE/ENTER key.
bAUd1= (Serial Port 1 Baud Rate)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER` key again to save it. Otherwise, use the /UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
12 = 1200 Baud 24 = 2400 Baud 48 = 4800 Baud
96 = 9600 Baud 14 = 14.4k Baud 19 = 19.2k Baud
28 = 28.8k Baud 38 = 38.4k Baud 57 = 57.6k Baud
76 = 76.8k Baud 11 = 115.2k Baud 23 = 230.4k Baud
Prty1= (Serial Port 1 Parity)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER` key again to save it. Otherwise, use the /UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
0 = NONE (No Parity) 1 = Odd Parity 2 = Even Parity
bitS1= (Serial Port 1 Data Bits)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER` key again to save it. Otherwise, use the /UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
7 = 7 Data Bits 8 = 8 Data Bits