20
SETUP AND CALIBRATION, Cont.
Hi rES - Display High Resolution Weight
With Hi rES on the display, pressing the
ENTER
key will show the active weight in "high resolution"
mode (in 1/10 interval). Press the
PRINT
key to print the weight (followed by the text TEST) via
the selected printer output port enabled during setup and calibration.
LoCoUt - Key Lock Out Function
With LoCoUt on the display, pressing
any
key will display "LoCd" (locked) or "UnLoCd" (unlocked)
for the current key state. Pressing a locked key during normal operation will results in a 1/2
second display "LoCd" and the key will be ignored. To exit the LoCoUt function, press the
calibration switch or cycle power (press the
ON/OFF
key twice).
NOTE! The menu selections Hi rES and LoCoUt can only be selected using the
calibration switch.
S E t U P
USA (domestic or international)
With SEtUP displayed, press the
ENTER
key. The display will change to USA=. Press the
ENTER
key to show the current value. If the setting displayed is acceptable, press the
ENTER
key
again to save it. Otherwise, using the numeric keys,
0/NO
or
1/YES
, enter the new setting, then
press the
ENTER
key to save it.
USA = 1 (Domestic) USA = 0 (International)
Date = mm/dd/yy Date = dd/mm/yy
Trl = no Trl = yes
Cap + 4% to OC Cap + 9 grads to OC
PT printed with tare
Lamptest on power up
LFt (Legal For Trade)
Press the
ENTER
key to show the current value. If the setting displayed is acceptable, press the
ENTER
key again to save it. Otherwise, using the numeric keys,
0/NO
or
1/YES
, enter the new
setting, then press the
ENTER
key to save it.
LFt = 1 LFt = 0
Interval Settings (Int=) allowed
are: 1, 2, 5, 10 , 20, 50
Interval Setting (Int=) is
selectable from 1 to 99.
NOTE!
When both
LFt=1 and USA=1
, the followings results occur:
Scale must have between 100 and 10,000 divisions
Tra = .5 or 0 to 3
Inhibit serial data during input
Disables
COUNT
key
Date = mm/dd/yy
Trl = no
Cap + 4% to OC
NOTE!
When
LFt=1 and USA=0
, the followings results occur:
Uns = 1
Date = dd/mm/yy
Trl = yes
Cap + 9 grads to OC
PT printed with tare
Lamptest on power up