Alaris System DFU – up to v9.12 Model 8015 General Information 3-53
PCA Module Section
Specifications and Symbols
Specifications
Bolus Dose Range: Configured according to hospital best-practice guidelines.
Bolus Volume, Maximum after
Occlusion:
Maximum Bolus Volume specifications are based on following standard
operating conditions:
Atmospheric Pressure: 645 - 795 mmHg
Disposable Type: #30883
Humidity: 20 - 90%
Rate: 5 mL/h
Syringe Type: BD 50/60 mL
Temperature: 68 ±4°F
Volume Collection Time: approximately 2 minutes
Critical Volume: Maximum over-infusion which can occur in the event of a single-fault
condition will not exceed 2% of nominal syringe fill volume during loading
and 1% of maximum syringe travel after syringe loading.
Delivery Units: mcg, mcg/h, mg, mg/h, mL, mL/h
Dimensions: 4.5" W x 15.0" H x 7.5" D (exclusive of security door)
Environmental Conditions:
Equipment Orientation: To ensure proper operation, Alaris System must remain in an upright
position.
Occlusion Pressure Limit Bolus Volume (mL)
Low
0.994
High
0.396
Operating Storage/Transport
Atmospheric Pressure 525 - 4560 mmHg
(700 - 6080 hPa)
375 - 760 mmHg
(500 - 1013 hPa)
Relative Humidity
(Avoid prolonged
exposure to relative
humidity >85%)
20 - 90%
Noncondensing
5 - 85%
Noncondensing
Temperature Range 41 - 104°F
(5 - 40°C)
-4 - 140°F
(-20 - 60°C)