Adding a Modality and Configuring Modality Image Quality
2010-11-04 2G8647 4-5
3. Select the best image and make note of the TFT Set/Contrast Table
for that image.
4. If you are satisfied with the best image, do the following.
a.
Exit the test screen.
b. On the Modality Configuration screen, select the Contrast Table and
TFT Set you want to use.
c. Touch Save.
5. If you determine that the best image is an improvement and you
want to continue fine tuning, do the following.
a.
Exit the test screen.
b. On the Modality Configuration screen, change the Contrast Table and
TFT Set you want to use.
c. Touch Contrast Test.
d. Run a Custom Contrast Test only (see “Performing a Contrast Test”
on page 4-6).
e. Touch OK.
f. Repeat step 5 until you determine the best image.
6. Backup the configuration. Touch:
Close
Close
Backup
Ye s
OK