Carestream Dryview 6850 - Environmental and Operational Requirements

Carestream Dryview 6850
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
2010-11-04 2G8647 6-3
Environmental Requirements
Temperature Operating: 15° to 35°C (59° to 95°F)
Storage: -35° to 60°C (-31° to 140°F)
Relative Humidity Operating: 20% to 85% RH, noncondensing
Storage: 10% to 90% RH, noncondensing
Altitude 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
Floor Vibration For the Imager to operate with no image degradation, floor vibration
levels cannot exceed the following.
Floor Levelness The floor must be level within 1°.
G
2
/Hz Freq. Range Direction
1 x 10
-7
1-50 Hz Vertical or horizontal
6 x 10
-5
50-200 Hz Vertical or horizontal
5 x 10
-6
200-650 Hz Vertical or horizontal
1 x 10
-6
650-1000 Hz Vertical or horizontal