22
CAL 2 (Minimum division calibration)
Function: Minimum division and decimal position setting
Range of set value: 0.001 ~ 50
Used key
Display
Descriptions
Using numeric keys ~
Enter minimum division
= Set, =Cancel
Push when entering a decimal
print
d= 0.001
Minimum division 0.001 kg
d= 0.01
Minimum division 0.01 kg
d= 0.1
Minimum division 0.1 kg
d= 1
Minimum division 1 kg
d= 10
Minimum division 10 kg
Reference 1. When CLEAR is pressed with decimal point calibrated, 'CAL2' is exited.
Reference 2. For external exploded view, set within the value of maximum capacity divided by the readability,
'1/30,000'. If the external exploded view is greater than '1/30,000', "Err20" will be displayed.
Reference 3. The decimal point positions is determined as the decimal position of the readability set from 'CAL2'.