7
- Series 1500 & 3000 Back Panel -
*1) Vent Probe Connector -
The vent thermocouple plugs in here.
2) Printer Port -
Connect a laser printer here for direct printout of activity log.
*3) Vent Solenoid Valve Connector -
The vent solenoid valve plugs in here.
4) Computer Port -
A computer terminal may be connected here for remote programming
and monitoring.
5) Sensor Hose Connector -
This port connects the sensor hose from the vessel.
6) Lid Switch Connector -
The lid monitor switch plugs in here and gives a text message on the
LCD screen whenever the lid is opened. This information is also stored in memory and will
appear in the activity log.
7) Lid Probe Connector -
The lid temperature thermocouple plugs in here.
8) Fill Solenoid Valve Connector -
The fill solenoid valve plugs in here.
9) Level Alarm Connector -
This port is used to attach an Auto-Dialer or a Remote Alarm
Monitor; This connector will give high and low level alarms, as well as power-loss alarm if
power to the unit is cut off.
10) Fuse -
1 amp Buss fuse; 220 volts slow-blow.
11) Power Cord -
This is the main power supply cord to the unit, be sure to check for proper
voltage - See voltage requirements listed on the controller.
IMPORTANT: Numbers 1 and 3 are installed only on units equipped with the Bypass Option
(24V .63 amps max)