Questions and Answers

  • J
    James HowellAug 17, 2025
    How to fix HIGH ALARM flashing on CBS V3000?
    • A
      Angela MaldonadoAug 17, 2025
      If the HIGH ALARM text is flashing and there's an audible alarm on your CBS Storage, it means the liquid level in the vessel is too high. First, push the STOP switch. If it continues to fill, the solenoid valve may be stuck open. Try pushing the FILL and STOP switches 15 to 20 times to unstick it. If that doesn't work, turn off the manual valve on the liquid nitrogen supply tank and let the solenoid valve warm up. Then, reopen the manual valve to see if it works. If not, the solenoid valve might need to be disassembled for service. Alternatively, there may be an obstruction in the solenoid valve. In that case, remove the solenoid valve and disassemble it for service.
  • J
    Jennifer WhiteAug 20, 2025
    Why is the LOW ALARM flashing on my CBS V3000?
    • A
      Antonio SmithAug 20, 2025
      If the LOW ALARM text is flashing and there's an audible alarm on your CBS Storage, it could be due to several reasons: * The supply tank is empty: Replace the supply tank. * The supply tank is shut off: Open the manual valve on the supply tank. * The supply tank pressure is too low: Check the pressure in the supply tank. It should be between 18 and 22 p.s.i. * Three or more storage units are filling at the same time from a single supply line: The storage vessels will fill slowly, and the alarm should correct itself.
  • M
    Mr. Michael Martin MDAug 23, 2025
    What to do if the SOURCE ALARM is flashing on CBS V3000 Storage?
    • Z
      zmckenzieAug 23, 2025
      If the SOURCE ALARM text is flashing and there's an audible alarm on your CBS Storage, it could mean that: * The supply tank is empty: Replace the supply tank. * The supply tank is shut off: Open the manual valve on the supply tank if it's closed. * The supply tank pressure is too low: Check the pressure; it should be between 18 and 22 p.s.i. * Three or more storage units are filling at one time from a single supply line: The storage vessels will fill very slowly, and the alarm will correct itself.
  • T
    taylor91Aug 26, 2025
    Why is the alarm silent on my CBS Storage?
    • J
      Julia WagnerAug 26, 2025
      If any ALARM text is flashing on your CBS Storage, but there is no audible alarm, it means the alarm MUTE is activated. To fix this, correct the alarm condition and then push the SOUND switch. This will activate the audible alarm for the next alarm occurrence.
  • M
    munozcolleenAug 28, 2025
    How do I resolve a TEMPERATURE ALARM on CBS V3000?
    • S
      Steven WangAug 28, 2025
      If the TEMPERATURE ALARM text is flashing and there is an audible alarm on your CBS Storage, this could be because the temperature has risen above the setpoint, possibly because the lid has been open too long. Alternatively, it could be due to a low liquid nitrogen level, in which case you should move the probe to its original location.
  • M
    michael21Aug 31, 2025
    What does it mean when LOW ALARM flashes and the level display reads '00.0 CM' on CBS Storage?
    • W
      wolfnancyAug 31, 2025
      If the LOW ALARM text is flashing, the audible alarm is present, and the Level Display reads '00.0 CM' on your CBS Storage, it could mean that the LN2 level in the vessel has fallen below the pressure sensing port. In this case, an automatic fill will begin, and the alarm will stop when the liquid level rises above the LOW level setpoint and the pressure port. Another possibility is that the pressure sensing hose is loose or disconnected. Check the pressure hose connections at both ends, ensuring a white clamp is in place at each end of the hose. If the hose seems loose, disconnect it, remove an inch off the end, and reconnect it.
  • R
    Reginald CabreraSep 3, 2025
    What causes the FILL VALVE alarm on my CBS V3000 Storage?
    • K
      Kenneth PalmerSep 3, 2025
      If the FILL VALVE text is flashing and there's an audible alarm on your CBS Storage, it could be because the fill valve is disconnected from the power source. Ensure the fill valve power cord is correctly connected. If the fill valve is correctly connected, the fill valve may be defective and need to be replaced.
  • S
    stephaniewatsonSep 5, 2025
    Why is the OPEN LID PROBE alarm flashing on my CBS Storage?
    • M
      mclaughlincassandraSep 5, 2025
      If the OPEN LID PROBE text is flashing and there's an audible alarm on your CBS Storage, it could be because the thermocouple probe plug is disconnected. Connect the probe plug properly. If the probe is connected properly, the thermocouple probe may be damaged and need to be replaced.
  • V
    Valerie EdwardsSep 8, 2025
    What does the OPEN VENT PROBE alarm mean on CBS V3000 Storage?
    • M
      Michael Flynn DDSSep 8, 2025
      If the OPEN VENT PROBE text is flashing and there's an audible alarm on your CBS Storage, it could be because the thermocouple probe plug is disconnected. Connect the probe plug properly. If the probe is connected properly, the thermocouple probe may be damaged and need to be replaced.
  • P
    Paul SheltonSep 12, 2025
    Why are there no lights on my CBS Storage unit?
    • M
      melanie31Sep 12, 2025
      If there are no lights on your CBS Storage, it could be due to several reasons: * Blown fuse: Replace it with a one amp fuse. * No power at the wall outlet: Use a live power source. * AC power cord not plugged in: Plug the AC power cord into a wall socket. * ON/OFF switch in the OFF position: Move the power keyswitch to the ON position.

Summary

Liquid Nitrogen Safety

Handling Liquid Nitrogen

Guidelines for safely handling liquid nitrogen, including proper protective clothing and container use.

Initial Set-up

Operation

Optional CryoMonitor Software

Front Panel Controls

Series 1500 & 3000 Back Panel

Series 5000 Back Panel

Series S1500 & S3000 Connections

Series V1500 & V3000 Connections

Series S5000 Connections

Series V5000 Connections

V Series Manual Fill Instructions

Remote Alarm Wiring

Storage Unit Parameters

LCD Display Functions

Standard Operating Mode Displays

Description of various display windows that appear in the Standard Operating Mode.

Program Mode Displays

Description of various display windows that appear when the unit is in Program Mode.

Series 1500 & 3000 Controller Wiring

S Series Liquid Level Display Verification & Calibration

S Series Level Display Verification

Steps to verify the accuracy of the S Series liquid level display against a yardstick.

V Series Liquid Level Display Verification & Calibration

V & S Series High & Low Level Alarm, and Temperature Display Verification

Troubleshooting Guide: Alarms

Troubleshooting Guide