OPERATOR’S MANUAL
4. CLEANING AND DISINFECTION
The outer parts of the equipment must be cleaned and disinfected using a product for hospital use with
indicationsforHIV,HBVandtuberculocidal(medium-leveldisinfectant)specicforsmallsurfaces.
The various drugs and chemical products used in dental surgeries may damage the painted surfaces and
the plastic parts. Researches and tests performed show that the surfaces cannot be fully protected against
theharshactionofallproductsavailableonthemarket.Wethereforerecommendprotectingwithbarriers
whenever possible.
The harsh actions of chemical products also depend on the amount of time they are left on the surfaces.
It is therefore important not to leave the product on the surfaces longer than the time specied by the
manufacturer.
If you do not use STER 1 PLUS, it is recommended to use products that contain at maximum:
• Ethanol at 96%. Concentration: maximum 30 g per 100 g of disinfectant.
• 1-Propanol (n-propanol, propyl alcohol, n-propyl alcohol). Concentration: maximum 20 g per 100 g of
disinfectant.
• Combination of ethanol and 1-propanol. Concentration: the combination of the two should be maximum
40 g per 100 g of disinfectant.
In particular, STER 1 PLUS (CEFLA S.C.) is recommended for daily cleaning operations.
WARNING!
• Do not use products containing isopropyl alcohol (2-propanol, iso-propanol).
• Do not use products that contain sodium hypochlorite (bleach).
• Do not use cleaners that contain phenols.
• All products must be used as directed by the manufacturer.
• Do not mix the STER 1 PLUS disinfectant with other products.
4.1.
CLEANING AND DISINFECTION INSTRUCTIONS
WARNING!
Carry out these operations with the light o (see paragraph 3.1) and after it has cooled down.
For cleaning and disinfection, use soft, non-abrasive disposable paper (avoid using recycled paper), or sterile
gauze.
Do not use sponges or, in any case, any material that can be reused.
• Light arms and optical unit cover.
Clean the outer parts of the equipment using STER 1 PLUS (CEFLA S.C.) or a similar product.
• Transparent screen and reflector.
Cleanusingasoftclothforglasssoakedinasolutionofneutralsoapandwater.
WARNING!
Never use products containing alcohol.
• Removable handles (VENUS PLUS, VENUS PLUS CL version, VENUS E version lights).
The handles can be cold-sterilized.
• Removable handles (VENUS PLUS -L version, VENUS PLUS CL -L version, VENUS PLUS MCT
version, VENUS PLUS CL MCT version lights).
Handles can be sterilized in a steam autoclave at 135°C (2 bar). Min. time 15 min.
A slight colour change of the handles is possible.
WARNING!
• It is strictly prohibited to use abrasive substances and detergents based on trichloroethylene,
gasoline, turpentine or solvents in general on any part of the light.
• Do not spray the selected product directly on the optical unit.
• All materials used to clean and disinfect must be thrown away.