14
CM73522UMC.0112
(6) [UP] Switch - increases the preset target volume by 5 ml each time pressed.
(7) [DOWN] Switch - decreases the preset target volume by 5 ml each time pressed.
(8) [
PRESET
] Switch - recalls one of the numbers that were factory-set. If the preset
numbers are 450 and 500 ml, the target volume displayed toggles as 500 - 450 -
500 - 450 - ... each time this switch is pressed. Power-on default figure is 500.
These preset numbers and default figure may be changed upon user’s request.
[The figure in Section 3-4. LCD Display Panel]
(1) Main Display - displays either total weight, blood volume or elapsed time. [ML],
[
GRAM
] or [
TIME
] is displayed to indicate what is being displayed.
(2) Preset Target Volume Display - has three digits. This figure is controlled by [UP],
[DOWN] or [PRESET] switch.
(3) Low Battery Indicator - is illuminated when battery requires recharging. Beep
tone sounds when this symbol is first activated.
(4) [ALARM] indicator - is illuminated when an abnormal condition occurs, such as a
slow draw. Refer to the Section 5-3 for details.
(5) [TIME] indicator - is illuminated only when the [TIME] switch is depressed. The
main display shows the elapsed time in [min : sec].
(6) [BLEED] indicator - is illuminated while blood collection is in progress. When the
[START] button is depressed, this indicator is on and [ML] is on too. This indicator
goes off and [GRAM] is on when the [END] button is depressed.
(7) [END] indicator - is illuminated when the collection is completed. When the
collected volume reaches the preset target volume, this indicator is illuminated.
The tube clamp closes and beeps sound. The front LED lamp also flashes.
(8) [ML] indicator - is illuminated when the [START] button is depressed. It means the
main display is now showing ‘net’ volume of blood collected in the blood bag in
milliliters.
(9) [GRAM] indicator - is illuminated when the machine is not in collection process. It
means the main display is now showing total weight loaded on the tray in grams.
[The figure in Section 3-5. Tube Clamp Module]
(1) Tube Clamp - presses blood bag tubing to stop blood flow. This clamp is opened
and closed by [CLAMP] switch except at the end of a normal collection. The
[END] switch must be pressed to indicate completion of the collection cycle, then
[CLAMP] switch is enabled to work. The clamp is automatically opened when the
[START] switch is pressed to start collection. It is automatically closed at the end
of collection, or when the [END] switch is pressed.
(2) Guide Pins - keep the blood bag tubing in proper position.
(3) Connection Plug - is a circular connector.
(4) Screws - should be removed to move the module to the other side.