POWER
Input 100 - 240 V AC, 1.0 to 0.42 A, 50/60 Hz
Electrical Class CLASS II
Mode of Operation Continuous
ELECTRICAL TYPE (DEGREE OF PROTECTION)
Ultrasound TYPE BF
Electrotherapy TYPE BF
Electrotherapy Vacuum TYPE BF
Ultrasound & Electrotherapy TYPE BF
GENERAL SYSTEM OPERATING AND STORAGE
TEMPERATURE
Operating Conditions
The device will meet its requirement under the following
conditions:
Temperature: 5°C to 40°C
Relative Humidity: 15 % to 90 %
Atmospheric Pressure: 70 kPa to 106 kPa
Transport and Storage Conditions
The device will remain in proper condition under the
following conditions:
Temperature: -20°C to 60°C
Relative Humidity: 10 % to 100 %
Atmospheric Pressure: 50 kPa to 106 kPa
Time required for the ME EQUIPMENT to warm from
the minimum storage temperature between uses until
the ME EQUIPMENT is ready for its INTENDED USE
when the ambient temperature is 20 °C: 5h
Time required for the ME EQUIPMENT to cool from the
maximum storage temperature between uses until the
ME EQUIPMENT is ready for its INTENDED USE when
the ambient temperature is 20 °C : 5h
IPXX RATING FOR UNIT
Rated to
IP21
IP2*
Protection against fingers or other object not greater
than 80mm in length and 12mm in diameter
IP*1
Protection from vertically dripping water
IPXX RATING FOR US APPLICATOR
Rated to IPX7
IPX7 Protection from immersed in water (up to 1m depth)
RED
RF transmitter/receiver characteristics:
Frequency Band transmission: 2400–2483.5 MHz
Modulation type: GFSK
Data rate: up to 2Mbps 500kHz deviation at 2Mbps
Effective radiated power: +6dBm