MAINTENANCE OPERATION
EN
| Intelect
®
Mobile 2 - Service Manual 24
4.4. ADVANCED SETTINGS
Persistent log storage:
Put to On to record the log continuously (use only for a short
time)
Can be used to diagnose and track a problem
Can be exported on USB key
Data transfer enabled:
Put to On to remove the Data transfer option
Clean Db:
Clean definitively the memory of deleted elements
Clean Master Db UUID:
Delete the identified computer when a synchronization with
Chattanooga application has been done
Ble Hw reset: Reset the Bluetooth Module if pairing issue
with computer
4.5 SERVICE MENU
On the “Display Unit Version Information” screen:
- Press and hold Play/Pause button
- Turn the knob 1 click to the right + 6 clicks to the left
- Release Play/Pause button
- Wait for a special Bip
- Return to Settings and find the menu: ServiceDiagnosis
On the “Display Unit Version Information” screen:
- Press and hold Play/Pause button ()
- Press the touch screen ( )
- Release Play/Pause button
- Wait for a “special Bip”
- Return to Settings and find the menu:
Advanced settings
-
Procedure P 29

Other manuals for Chattanooga Intelect Mobile 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chattanooga Intelect Mobile 2 and is the answer not in the manual?

Chattanooga Intelect Mobile 2 Specifications

General IconGeneral
BrandChattanooga
ModelIntelect Mobile 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals