Local Operation
4. Set Slew Rate
There are 250 discrete steps in each range. Set the rise 20 A/µS and fall slew rates to
0.5A/µS by pressing 2, 0, ENTER for rise and 0, . , 5 ENTER for fall slew rate.
CCH : 20.0A/uS
CCH : 0.5A/uS
5. Set Voltage spec
Set the high voltage spec. to 6 V by pressing 6 , ENTER.
CCH VOLTAGE SPEC.
HIGH : 6.000V
Set the low voltage spec. to 4 V by pressing 4 , ENTER.
CCH VOLTAGE SPEC.
LOW : 4.000V
Then the display will go to the first editing page again.
L Notice
Spec. value can also be programmed by PERCENT under Configure setting. You have to
input the CENTER Voltage, and then enter the HIGH and LOW percent to the center voltage.
6. Set dynamic current mode.
Press DYNA key under CC mode, and select current range by RANGE key. Program
the high current level to 50A by pressing 5, 0, ENTER keys. Program the low current
level to 5A by pressing 5, ENTER keys.
CCDL1: 50.000A
CCDL2: 5.000A
Then set the dynamic function period T1 to 0.1 mS, period T2 to 0.2 mS by pressing
0, . , 1, ENTER and 0, . , 2 , ENTER . The range of dynamic period is from 0.025
mS to 30 Sec.
4-5