7.2.4 FETch Sub-system
FETch
:CURRent
:POWer
:RESistance
:VOLTAGE
:STATus
FETCh:CURRent?
Description Return the current measured at the input of the load.
Query Syntax FETCh:CURRent?
Return Parameters <NR2> [Unit=Ampere]
Query Example FETC:CURR?
Return Example 3.15
FETCh:POWer?
Description Return the power measured at the input of the load.
Query Syntax FETCh:POWer?
Return Parameters <NR2> [Unit=Watt]
Query Example FETC:POW?
Return Example 3.15
FETCh:RESistance?
Description Return the resistance measured at the input of the load
Query Syntax FETCh:RESistance?
Return Parameters <NR2> [Unit=OHM]
Query Example FETC:RSE?
Return Example 10.22
FETCh:VOLTage?
Description Return the voltage measured at the input of the load.
Query Syntax FETCh:VOLTage?
Return Parameters <NR2> [Unit=Voltage]
Query Example FETC:VOLT?
Return Example 8.12
FETCh:STATus?
Description Return the real time status of the load.
Query Syntax FETCh:STATus?
Return Parameters <NR1>