High Power DC Electronic Load 63200 Series Operation & Programming Manual
load.
Return Example 2.5
CURRent:DYNamic:T1/T2
Description Set the duration parameter T1 or T2 for dynamic load.
Syntax CURRent:DYNamic:T1 <NRf+>[suffix]
CURRent:DYNamic:T2 <NRf+>[suffix]
Parameters For valid value range refer to respective specification.
Example CURR:DYN:T1 10mS Set dynamic duration T1 = 10mS.
CURR:DYN:T2 2S Set dynamic duration T2 = 2S.
CURR:DYN:T1 MAX Set dynamic duration T1 to the
maximum value.
CURR:DYN:T2 MIN Set dynamic duration T2 to the
minimum value.
Query Syntax CURRent:DYNamic:T1?
CURRent:DYNamic:T2?
CURRent:DYNamic:T1? MAX
CURRent:DYNamic:T2? MIN
Return Parameters <NR2> [Unit=Sec]
Query Example CURR:DYN:T1? Return the dynamic duration parameter
T1.
Return Example 0.15
7-12