EasyManua.ls Logo

Chroma 63202 - Page 89

Default Icon
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Language Dictionary
CONF:VOLT:LAT OFF Set the action type toVon as Non Latch
(Refer to the former part of this
manual for detailed).
Query Syntax CONFigure:VOLTage:LATCh?
Return Parameters 0 | 1
Query Example CONF:VOLT:LAT? Return the action type to Von.
Return Example 0 (OFF)
CONFigure:VOLTage:LATCh:RESet
Description Reset Von signal.
Syntax CONFigure:VOLTage:LATCh:RESet
Parameters None
Example CONF:VOLT:LATC:RES
CONFigure:VOLTage:ON
Description Set the voltage of sink current to on.
Syntax CONFigure:VOLTage:ON <NRf> [suffix]
Parameters For valid voltage range refer to respective specification.
Example CONF:VOLT:ON 1 Set Von=1V.
CONF:VOLT:ON 300mV Set Von=300mV.
Query Syntax CONFigure:VOLTage:ON?
Return Parameters <NR2> [Unit=Voltage]
Query Example CONF:VOLT:ON? Return the setting of Von value.
Return Example 3.5
CONFigure:VOLTage:POLarity
Description Set the voltage display polarity.
Syntax CONFigure:POLarity <NR1 | CHAR>
Parameters 0 | MINUS, 1 | PLUS,
Example CONF:POL 0 Set voltage display to minus.
CONF:POL PLUS Set voltage display to plus.
Query Syntax CONFigure:POLarity?
Return Parameters 0 | 1
Query Example CONF:POL? Return the display polarity.
Return Example 0 ( MINUS )
CONFigure:VOLTage:RANGe
Description Set voltage measurement range in CC mode.
Syntax CONFigure:VOLTage:RANGEe <NR1 | CHAR>
Parameters 0 | L, 1 | H
Example CONF:VOLT:RANG 1 Set full-range to High.
CONF:VOLT:RANG L Set full-range to Low.
Query Syntax CONFigure:VOLTage:RANGe?
Return Parameters 0 | 1
Query Example CONF:VOLT:RANG? Return Voltage range.
Return Example 1 ( H )