SYSTEM SPECIFICATIONS
CO2 Analyzer:
Sensor: Low power Non-Dispersive Infrared Gas Analyzer. No sensitivity to motion
Chopping frequency: 1 Hz
Source life: 5000 hours
Sensor response time: 35 seconds
Repeatability: ± 0.1ppm (short term)
Sample cell: 100 mm × 10.2 mm (3.94”L × 0.40” Dia)
Warm-up time: Approximately 3 minutes
Accuracy: < ± 2% up to 2000 ppm
Resolution: 0.1 ppm.
Power supply: 7.2 VDC, 4400 mAh for 5 hours continuous use, extended hours of use with additional
batteries. AC Adapter / Battery Charger supplied
Power consumption: 2.5 watts (sample pump in operation)
Pump flow rate: 100 ~ 1000 cm2 / min
Mass flow meter Accuracy: 2%
Display: LCD 40 x 6 characters or 320 x 64 pixel
Data output: PC link cable, RS232 or USB (with adapter)
Data storage: 4 MB internal FLASH RAM
Operation temperature: 0 to 45ºC, 0 - 90% RH non-condensing
Dimensions: 44 cm x 5.5 cm x 5 cm
Weight: 1.5 Kg., (3 lb.) (with battery)
Standard range: 0 to 2000 ppm (Standard) - 0 to 3000 ppm (Optional)
H2O Analyzer
Sensor Type: Humidity Sensitive Capacitor
Stability: Stable Analyzer for Accurate H20 Measurements
Measuring Range: 0 to 100% RH
Resolution: 0.1%
Accuracy: ±2% at 10% RH, ±3.5% at 95% RH
Sensor response time: 15 seconds