Elektrische Daten
DC Voltage
Range Resolution Accuracy Overload protection
200 mV 100 μV
± (0,5 % + 2)
250 V DC or AC
2000 mV 1 mV
20 V 10 mV
200 V 100 mV
250 V 1 V ± (0,8 % + 2)
AC Voltage
Range Resolution Accuracy Overload protection
200 V 100 mV ± (1,2 % + 3) 250 V DC or AC
250 V 1 V
DC Current
Range Resolution Accuracy Overload protection
200 μA 0,1 μA ----------- 200 mA 250 V, F (fast), 5×20 mm
2000 μA 1 μA ± (1 % + 2) 200 mA 250 V, F (fast), 5×20 mm
20 mA 10 μA ± (1 % + 2) 200 mA 250 V, F (fast), 5×20 mm
200 mA 100 μA ± (1,2 % + 2) 200 mA 250 V, F (fast), 5×20 mm
10 A 10 mA ± (2 % + 5) 10 A, 600 V, F (fast), 6×25 mm
Remark: At 10A range: For continuous measurement ≤ 10 seconds and interval not
less than 15 minutes.
Resistance
Range Resolution Accuracy Overload protection
200 Ω 0,1 Ω ± (0,8 % + 5) 250 V DC or AC
2000 Ω 1 Ω ± (0,8 % + 2)
20 kΩ 10 Ω ± (0,8 % + 2)
200 kΩ 100 Ω ± (0,8 % + 2)
20 MΩ 10 kΩ ± (1 % + 5)
200 MΩ 100 kΩ ± (5 % (reading -10) + 10)