CleanSpace Respirators USER INSTRUCTIONS
19
With the mask off your face, pull the
orange plug from the Sensing Ball
socket.
Locate the Sensing Ball which
was supplied with your mask.
Fit the Sensing Ball. Push it all the way in.
Note that there are small fins on the stem of
the ball to ensure the correct orientation.
Remove the sealing plug from
the fit test port.
Fit the mask sensing tube to the fit test port
(on a TSI PortaCount it is the clear tube).
Re-fit the inner mask. The inner mask is an essential part of the Full Face Mask and is required to complete the fit test. Don the
mask and begin the test, following the protocol provided with the fit testing machine. A CleanSpace Respirator with a clean
filter, a full battery, and the Power Unit on, should achieve a fit factor above 1000. If your result is below 1000, adjust the
mask and try again.
WARNING
If a satisfactory quantitative fit factor or a qualitative test pass cannot be achieved with either mask, the
CleanSpace Respirator must not be used.
Satisfactory quantitative fit factor results are as follows.
On completion of the fit test, the Full Face Mask must be prepared for use as follows.
Remove the Sensing Ball from the
Sensing Ball socket. Refit the orange
plug to the Sensing Ball Socket,
pushing it all the way in.
Remove the sensing tube from the fit
test port. Refit the plug to the fit test
port.