9
INTOXILYZER
®
S-D2 Operator’s Manual
checking to insure the fuel cell is discharged and free of
alcohol from any previous sample. Elimination of alcohol
from the fuel cell should take no longer than two
minutes—except in unusual cases—depending on how
much alcohol was actually present in the last sample.
To conduct a READY CHECK, depress the ‘READ’ button
and hold it down for at least ten seconds. This
will release the ‘SET’ button and switch on the amplifier
and display circuits. Observe the display; it should not
exceed .002 after ten seconds. If the display does not
show .002 or less during the ten second READY CHECK,
the fuel cell may have traces of residual alcohol from a
previous sample. If this occurs after a previous test,
depress the ‘SET’ button to lock and wait one minute
before repeating the READY CHECK. If the left digit
shows ‘L, replace the battery.
2) SET
Once the READY CHECK is complete, the sampling
system must now be SET to prepare it to draw a breath
sample into the fuel cell for analysis. Press down the
‘SET’ button until it locks. This pushes down the
diaphragm and holds it against a spring-loaded catch.
This action also places a short-circuit across the fuel cell,
which accelerates its inter-sample recovery time.
To complete the READY
CHECK, depress the
‘READ’ button for at least
ten seconds.