26
INTOXILYZER
®
S-D2 Operators Manual
If the display reading cannot be brought up to the
dry gas standard value, the fuel cell has reached the
end of its life and should be replaced. See the next
section for more information.
6) RESET AND WAIT
After the final alcohol reading is noted, depress the ‘SET’
button to lock. This allows the fuel cell to clear itself of
alcohol before using the instrument for more tests. This
will take approximately two minutes.