EasyManua.ls Logo

CMi Intoxilyzer S-D2 - Breath Sampling Light Check

CMi Intoxilyzer S-D2
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
INTOXILYZER
®
S-D2 Operator’s Manual
ITS UP POSITION. THIS WOULD RUPTURE THE
SAMPLING DIAPHRAGM OR FUEL CELL
ELECTRODE AND NECESSITATE THE
REPLACEMENT OF EITHER OR BOTH OF THESE
PARTS.
BREATH SAMPLING LIGHT CHECK
This procedure checks the operation of the breath
pressure switch and the timing and operation of the
breath sampling lights:
Check the
S-D2’s sampling
system by 1)
pressing the
‘SET’ button to
the locked
position and
placing a
forefinger over
the sampling
port. 2) Press
the ‘READ’
button down
and observe the
‘SET’ button,
which should
not rise.

Related product manuals