INTOXILYZER
®
S-D2 Operators Manual
CALIBRATION ADJUSTMENT
CHECKLIST
1. Preliminary
2. Ready Check
3. Set
4. Sample the Dry Gas Standard
5. Adjust the Display to the Value of
the Dry Gas Standard
6. Reset and Wait
DETAILED PROCEDURE
The calibration adjustment process assumes three
conditions:
• The instrument has not analyzed more than
two samples containing alcohol within the
previous hour or any sample within the
previous ten minutes,
• The instrument is in its normal operating
temperature range, and
• The battery does not need replacement.
1) PRELIMINARY
The instrument must first be removed from its protective
leather pouch. Remove the wrist strap by detaching the
split ring from the metal pillar. Then, insert the
screwdriver included with the S-D2 into the hole in the
pouch and ease the leather upwards taking care to
prevent the pillar from catching on the pouch. At the
same time, push the instrument upwards from the
bottom of the pouch, easing the pillar under the edges of
the hole.