24 NUCLEUS 7 SOUND PROCESSOR USER GUIDE
Use monitor earphones
Carers can use monitor earphones to check that a Cochlear
implant recipient is receiving sound, and that functions like
telecoil or wireless accessories are working.
NOTES
Only use approved earphones/headphones with the monitor
earphone adaptor (ones with a 3- or 4-pin audio jack that
fits the adaptor). You CAN use earbuds. You CANNOT use
noise-cancelling headphones.
Don’t connect multiple monitor earphone adaptors in series.
Don’t connect the monitor earphone adaptor to other audio
outputs, e.g. computer.
1. Twist to disconnect, then
remove the battery
module.
2. Insert the Cochlear
Nucleus 7 Monitor
Earphone Adaptor, then
twist to connect.
DRAFT ONLY FOR CLINICAL TESTING
592753_1-06_EN_CP1000_UG_EMEA.indd 24 14/06/2016 1:23:12 PM