EasyManuals Logo

COMAC CAL FLOW 38 User Manual

COMAC CAL FLOW 38
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
Manual FLOW 38 Page 45 (total 54) COMAC CAL s.r.o.
Factory settings
The current loop is set in such a manner that 4 mA corresponds to zero flow and 20 mA corresponds to
its maximum value.
The address of the meter is set to 1 by default and communication parameters to 2400Bd,8db,1sb,parity
EVEN (Mbus) or 9600Bd,8db,1sb, no parity (Modbus).
Access password (PIN) for parameter changing is always set to 0000 and this value will be reset in case
of restoring to factory default settings.
Impulse constants and current loop factory settings
Diameter nominal
Impulse output
4 20mA (in Qmin/Qmax 1/100 range)
DN
Vout[imp/l]
Vout - pulse width [ms]
Q[l/h] for 4mA
Q[l/h] for 20mA
4
10
4
0
600
6
10
4
0
1,200
8
10
4
0
2,200
10
10
4
0
3,400
15
10
4
0
7,600
20
10
4
0
14,200
25
10
4
0
21,000
32
1
4
0
34,000
40
1
4
0
54,000
50
1
4
0
84,000
65
1
4
0
144,000
80
1
4
0
220,000
100
0.1
4
0
340,000
125
0.1
4
0
534,000
150
0.1
4
0
760,000
200
0.1
4
0
1,350,000
300
0.1
4
0
3,052,000
400
0.1
2.5
0
5,400,000
500
0.1
2.5
0
8,480,000
600
0.01
2.5
0
12,200,000
Diameter
nominal
Resolution V
Resolution Q
DN≤15
V [0.001 m3]
Q [0.001 m3/h]
50≥DN>15
V [0.01 m3]
Q [0.01 m3/h]
DN>50
V [0.1 m3]
Q [0.1 m3/h]

Other manuals for COMAC CAL FLOW 38

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the COMAC CAL FLOW 38 and is the answer not in the manual?

COMAC CAL FLOW 38 Specifications

General IconGeneral
BrandCOMAC CAL
ModelFLOW 38
CategoryMeasuring Instruments
LanguageEnglish