PRECAUTIONS WHEN USING ELECTRODES
CAUTION
Only use electrodes supplied by DJO, LLC� Other electrodes may have electrical properties that are
unsuitable for the Compex Wireless USA stimulator
• Do not use electrodes with a surface <16 cm² (2�94 in²), as there will be a risk of suffering a burn injury
Caution should always be exercised with current densities > 2mA/cm² (12�9 mA/in²)�
• Always turn off the stimulator before moving or removing any electrodes during a session�
• Do not place the electrodes in water
• Do not apply solvents of any kind to the electrodes�
• For best results, wash and clean the skin of any oil and dry it before attaching the electrodes�
• Attach the electrodes in such a way that their entire surface is in contact with the skin�
• For obvious reasons of hygiene, each user must have his/her own
electrode set� Do not use the same electrodes on different people�
Never use a set of adhesive electrodes for more than 15 sessions
as their bonding power deteriorates over time and optimal contact is very important for both user comfort
and efficacy
Some people with very sensitive skin may experience redness under the electrodes after a session�
Generally, this redness is completely harmless and disappears after 10 to 20 minutes� Never start another
stimulation session in the same area, however, if the redness is still visible�